line decor
You are here: home\applications\eSpira\parameters
line decor

 
 
 


eSpira™ Forced Manoeuvers System

FEV Parameters

Name

Units

Description

IC

ml

Inspiratory Capacity, volume inspired during slow inspiration

FVC

ml

Forced Vital Capacity, volume expired during fast expiration

ERV

ml

Expiratory Reserve Volume, FVC-IC

FEV25

ml

Volume expired in first 25 ms of fast expiration

FEV50

ml

Volume expired in first 50 ms of fast expiration

FEV75

ml

Volume expired in first 75 ms of fast expiration

FEV100

ml

Volume expired in first 100 ms of fast expiration

FEV200

ml

Volume expired in first 200 ms of fast expiration

FEV400

ml

Volume expired in first 400 ms of fast expiration

FEVPEF

ml

Volume expired at peak expiratory flow

tI

s

Inspiration time

tE

s

Expiration time

Pef

ml/s

Peak expiratory flow

MMEF

ml/s

Mean Mid Expiratory Flow, average flow between 25% - 75% FVC

FEF75

ml/s

Forced Expiratory Flow at 75 % of FVC (25 % expired)

FEF50

ml/s

Forced Expiratory Flow at 50 % of FVC

FEF25

ml/s

Forced Expiratory Flow at 25 % of FVC (75 % expired)

FEF10

ml/s

Forced Expiratory Flow at 10 % of FVC (90 % expired)

dVPEF

%

% of FVC remaining at PEF

HoldTime

s

Measured breath hold time

pInsp

cmH2O

Pressure difference between start and end of inspiration

pExp

cmH2O

Pressure difference between start and end of expiration

Offset

ml/s

If the auto-detect zero flow level is enabled, this outputs the offset from calibrated flow zero to calculated flow zero.  If zuto-zero detect is disabled, then this output is 0.

FEVUser1

ml

FEV at user-defined output point 1.

FEVUser2

ml

FEV at user-defined output point 2.

FEVUser3

ml

FEV at user-defined output point 3.

FEVUser4

ml

FEV at user-defined output point 4.

FEVUser5

ml

FEV at user-defined output point 5.

FRC Parameters

Name Units Description

FRC

ml

Functional Residual Capacity

DeltaP

cmH2O

Pressure difference

DeltaV

ml

Volume difference

Pb

mmHg

Barometric Pressure, as set in Control Options

Tol

%

Percent tolerance between accepted breaths

DeadSp

ml

Amount of dead space, as set in manoeuvre options

Quasi Static PressurePressure Volume Parameters

Name

Units

Description

IC

ml

Inspiratory Capacity, volume inspired during slow inspiration

FVC

ml

Forced Vital Capacity, volume expired during fast expiration

FRC

ml

Functional Residual Capacity, as entered in manoeuvre options

ERV

ml

Expiratory reserve volume, FVC - IC

TLC

ml

Total Lung Capacity, FRC + IC

RV

ml

Residual Volume

Pfrc

cmH2O

Pressure at FRC

Cmax

ml/cmH2O

Maximum compliance

P_Cmax

cmH2O

Pressure at max compliance

V_Cmax

ml

Volume at max compliance

Cchord

ml/cmH2O

Chord compliance between 0-10 cmH2O

Cfvc50

ml/cmH2O

Compliance at 50 % VC

C_P0

ml/cmH2O

Compliance at 0 pressure

Pmax

cmH2O

Max pressure

Pmin

cmH2O

Min pressure

tHold

s

Breath hold time

 